Related Products KC761A KC761B KCSDI

  1. Connect the KC761 with Ethernet cable.
  2. Press FUNC, communication, access the Ethernet Config menu on the KC761.
  3. Set the KC761’s network configuration to match the network segment of your PC. For example:
    • Local IP Config: If your PC’s IP is 192.168.1.100, set the KC761’s IP to 192.168.1.x(for example x = 101).
    • Local Gateway Config: Use the same gateway as your PC.
    • Local Subnet Mask: Use the same subnet mask as your PC.
    • Local TCP/IP Port Number: Ensure this is a unique port not used by other applications (default is typically work).
  4. Enable the Ethernet Switch on KC761.

Network Connection Scenarios

1. Direct Connection (PC to KC761)

  1. Connect the Ethernet cable directly between the PC and KC761.
  2. Set your PC as a Static IP mode
  3. Configure the KC761 IP settings to be on the same network segment as the PC.

2. Using a Network Switch or Router

  1. Connect both the PC and KC761 to the same network switch or router using Ethernet cables.
  2. Ensure both devices are powered on and connected to the network.
  3. Configure the KC761 IP settings to be on the same network segment as the PC.

Step 3: Connect KCSDI Software to KC761

  1. Ensure both the KC761 and PC are powered on and properly connected via Ethernet.
  2. Install and run the KCSDI software, add KC761 device and enter the KC761’s IP address and TCP/IP port.
  3. Click the ok button in the KCSDI software.

Troubleshooting

  • No Connection:
    • Verify that the Ethernet cable is securely connected.
    • Verify KC761 Ethernet Switch is ON
    • Ensure both devices are in the same network segment.
    • Double-check the IP address, subnet mask, and gateway settings.
    • Ensure no other devices on the network are using the same IP address.
Currency